Nidek YC1800 GYC1000 Combo Laser

Condition: Refurbished
Categories: ,
Contact for Price & Availability

Description

This unit has been completely inspected by our staff, all optics have been cleaned, software updated, and a service calibration has been performed.

All functions are working to the manufacturer’s specifications and this Combo YAG/Argon is ready for use. Please view all photos and feel free to ask questions!

Easily transition from YAG capsulotomies to 532nm Retina wavelength with this all-in-one platform!

YC-1800 Yag Laser:

The YC-1800 offers 0.3 to 10 mJ, continuously adjustable in increments of 0.1 mJ, allowing the precise tissue effect.

The dual 635 nm aiming beam offers superior visual sensitivity, which in turn enhances the speed and ease of operation. The 635 nm beam is also safer to human eyes since it achieves the same crisp and sharp intensity of 650 nm or 670 nm beam with half the power output. The YC-1800 has the ability to rotate the dual aiming beams 360º permitting work anywhere in the periphery without clipping the iris.

The YC-1800 can adjust the offset between ±500 microns (25 micron steps) to meet the varied clinical needs. A different offset can be used for PMMA, silicone or acrylic lenses, and the offset can even be adjusted on the same IOL to compensate for a parallax effect in the periphery. This eliminates the need to manually defocus, permits a clear field of view, and reduces lens pitting.

Specifications: 

-Type: Nd: YAG

-Wavelength: 1064 nm

-Mode Structure: Fundamental

-Pulse Duration: 4 nsec

-Mode of operation: Q-switched

-Pulse repetition rate: 3 Hz (single) / 1.5 Hz (burst)

-Output energy: 0.3 to 10.0 mJ / pulse (continuously variable)

-Burst mode: 2 or 3 pulse / trigger

-Spot size: 8 μm

-Cone angle: 16º

-Focus Shift: 0 to 500 μm (continuously variable, toward both anterior chamber and Posterior chamber

532nm Green Laser

Efficient, safe photocoagulation is a hallmark of the GYC- 1000. The 532nm laser beam passes through the ocular media with low attenuation to minimize power loss.

– Higher absorption by the pigment epitheliopathy, hemoglobin, and oxidized hemoglobin

– Lower absorption by the xanthophyll pigment

Exposure time of conventional lasers can be adjusted in 0.10 second increments, from 0.10 to 1.00 seconds. The GYC-1000’s exposure time can be adjusted in 0.05 second increments from 0.10 to 0.50 seconds, which is the most widely used range for photocoagulation. The finer adjustments provide more precise treatment for patients.

Specifications:

Treatment Laser: Frequency-Doubled Diode Pumped Solid State Laser

Wavelength: Green: 532nm

Output power: Green: 50-1700mW

Output type: Continuous Wave

Exposure times: 0.01-3.00 seconds

Automation repeat: 0.1-1.0 seconds intervals

Aiming laser: Red Diode

Power supply: 90-264Vac, 50/60Hz, 200VA

WE SHIP WORLDWIDE and have experience shipping to over 100 countries around the world.  Please email or message us for an international quote.

More photos available! Just ask.

Keep in mind we are not brokers. We are a full sales and service center. Each laser and piece of equipment we sell is hand picked, inspected and tested by our technicians and engineers to ensure your order arrives in excellent condition both cosmetically and mechanically. Buy with confidence from Laser Locators!

Financing

We offer easy and flexible financing options on new and used ophthalmic equipment, including, Bausch & Lomb, Visx, Zeiss Humphrey, Topcon, Canon, Alcon and more! Maximize your budget and improve your bottom line with Laser Locators easy equipment financing plan. We have partnered with Stearns Bank to provide financing to our customers.